The austenite finish temperature of the finished tube (active Af) with outer diameter in the range of 0.3 to 3.0 mm (0.0;
Basic Info.
Application Medical, Veterinary
Wall Thickness Range 0.05~0.1mm
Od Range 0.30~1.00mm
Sterilization Non Sterilization
Transport Package Box
Specification CUSTOMIZED
Trademark Peiertech
Origin China
Product Description

The austenite finish temperature of the finished tube (active Af) with outer diameter in the range of 0.3 to 3.0 mm (0.012 to 0.12 in.) shall be measured on representative full round tube sample(s) using the bend and free recovery method described in Test Method ASTM F 2082-15.


ASTM F2633-13 Wrought Seamless NiTi Superelastic Tube for Medical Device and ImplantsSuperelastic properties are measured per Test Method ASTM F2516 at room temperature (22+/-2C) Peiertech provides custom tubing manufacturing, including nitinol SE tube and SMA tube.
Outer Surface: Oxide or Center-less-ground
Inner Surface: Etched or Oxide
Other finished can be discussed with Peiertech.
Mechanical Properties
ConditionUTS
σb MPa
Elongation
δ%
Upper Plateau Stress
UPS MPa
Permanent set after 6% strain %Active Af
ºC
Superelastic≥1000≥10≥380≤0.3≤15

Permissible Variation in OD or ID Dimensions of Guide Tube*
Nominal OD (mm)OD Tolerance (mm)ID Tolerance (mm)
OD≤0.3±0.005±0.010
0.3<OD≤0.5±0.007±0.015
0.5<OD≤1.5±0.015±0.020
1.5<OD≤2.5±0.020±0.030
2.5<OD≤3.5±0.020±0.040
3.5<OD≤5.0±0.030±0.050
5.0<OD≤10.0±0.050±0.050

Permissible Variation in OD or ID Dimensions of Tube for Stents*
Nominal OD (mm)OD Tolerance (mm)WT Tolerance (mm)
0.3<OD≤0.6±0.015±0.015
0.6<OD≤1.5±0.020±0.015
1.5<OD≤2.5±0.020±0.020
2.5<OD≤3.5±0.020±0.020
3.5<OD≤5.0±0.030±0.025
5.0<OD≤10.0±0.050±0.030
10.0<OD±0.050±0.050